UV-A sources, filters, cameras and other equipment
equipment
illumination
filters
Author
Pedro J. Aphalo
Published
2022-10-15
Modified
2025-11-22
Abstract
This page contains notes about photography of ultraviolet-A-radiation induced visible fluorescence. I discuss methods in detail and provide examples of higher plants, lichens and mosses. I describe the ultraviolet-A sources, UV-pass and UV-blocking filters, lenses and cameras that I use, as well as camera settings I use.
Keywords
LED light, ultraviolet
Warning 1: Ultraviolet-A radiation dangers
Ultraviolet-A (UV-A) radiation at high irradiance can be damaging to eyes and skin, especially when there is little or no visible light together with it. Eye protection is recommended in all cases. It is safer to use only the so called “black light” corresponding to UV-A1 at wavelengths near 365 nm. However, even in this case when using flashlights or any other sources of UV-A that emits a concentrated beam, use of safety goggles is a must. When selecting goggles or safety spectacles pay attention to their UV-protection rating (e.g., UV400 CE, or increasing protection from ANSI U2 to U6 markings) as some goggles are certified to protect only from mechanical impacts (e.g., ANSI Z87+ with no U marking) (Figure 1). It is important that goggles and other safety eye wear provide protection also from the sides.
A: In white light.
B: In ultraviolet-A radiation.
Figure 1: Protection eye wear protects the eyes by blocking all UV-A radiation from reaching them. The ones in the photographs are simple ones, given away by an optician to me. They are anyway effective, looking black and casting a dark shadow on the white background. Photographs were taken with a full-spectrum converted Olympus E-M1 camera, using either no filter on the lens and Sunwayfoto FL-96 fill light video lamp or a Jaxman Uc1 UV-A flashlight, and filters ZWB2 stacked with TSN575 on the lens. Lens: Sigma 30 mm 1:2.8 DN MFT. The background was a 5 mm-thick slab of PTFE. A black reference is included in the top-right corner to facilitate consistent editing.
Viewing photographs
Clicking on photographs opens a gallery view with them displayed at a larger size and at higher resolution. This view also allows navigation among multiple photographs from a given Figure.
Folded code chunks
In this page code chunks can be “folded” so as to decrease the clutter. Above each plot, table or other R-code output you may find one or more “folded” code chunks signalled by a small triangle followed by “Code”. Clicking on the triangle “unfolds” folded chunks making the R code that produced the printed values, plots or tables visible. Clicking on the same icon on an “unfolded” chunk, folds it hiding the code.
The code when visible in the chunks can be copied by clicking on the top right corner, where an icon appears when the mouse cursor hovers over the code listing.
The </> Code drop down menu to the right of the page title makes it possible to unfold and fold all code chunks in the page and to view the Quarto source of the whole web page.
Fluorescence occurs when a molecule absorbs radiation and re-emits the absorbed energy also as radiation. The usual case is the capture of a single photon, dissipation of a (small) fraction of the energy as heat and the emission of the rest of the energy as a new photon. As the emitted photon carries less energy than the absorbed one, fluorescence takes place at a longer wavelength than the absorbed radiation.
Both UV and visible light can induce the fluorescence of chlorophyll in plants. Strong fluorescence is emitted at wavelengths near 685 nm and 740 nm, by chlorophyll-a and chlorophyl_b_. Phenolic acids fluoresce in the blue and green regions. Pigments in lichens, some mushrooms and even insects (as well as scorpions) fluoresce in a variety of bright pure colours.
We do not see the fluorescence in normal daylight as the high sensitivity of human vision with a peak in green prevent it. With a camera modified to increase sensitivity to infra-red radiation and a filter that blocks wavelengths shorter than 680–700 nm the fluorescence can be imaged. A flash of strong light is most effective and is used in the study of plant’s photosynthesis (Figure 2).
Figure 2: NIR chlorophyll fluorescence induced by a flash of white light in plants of fava beans. Camera: full-spectrum converted Olympus E-M1, Sigma 30 mm 1:2.8 DC DN MFT lens, Heliopan RG680 filter on lens. Strong white light illumination from a device used to study photosynthesis of plants.
The fluorescence spectrum can be measured, with some difficulty with an array spectrometer and a UV-A flashlight resulting in very noisy data.
Figure 3: Whole-leaf fluorescence spectrum from a lettuce leaf illuminated with UV-A1 radiation at 365 nm from a flashlight and measured on the opposite side of the leaf with a Maya 2000 Pro array spectromter with a cosine diffuser. (Heavily smoothed.)
Of course it can be also measured with more sofisticated equipment to obtain a clean fluorescence spectrum, in this case for a wheat leaf, measured under 355 nm excitation (Figure 4).
Figure 4: Whole-leaf fluorescence spectrum from a wheat leaf illuminated with UV-A1 radiation at 355 nm.
The quantum yield of fluorescence depends on the molecules involved. In most cases the fraction of the incident radiation that is emitted as fluorescence is small, frequently not even detectable. When we illuminate a plant or object with ultraviolet-A (UV-A) radiation, only some of its components, if any, fluoresce strongly, emitting visible light (VIS) and NIR radiation of specific wavelengths.
To induce fluorescence across the whole visible spectrum we have to use radiation of shorter wavelengths, such as UV-A radiation. Our eyes’ low sensitivity to UV-A radiation and easily available eyeglasses and goggles that block all UV radiation help in allowing us to see the fluorescence without it being overwhelmed by the radiation used as excitation.
Different lichens contain pigments that fluoresce brightly when exposed to UV-A radiation (Figure 5). Colours of the fluorescence vary between yellow, red, occasionally blue and rarely green.
A: In white light.
B: In ultraviolet-A radiation.
Figure 5: Lichens on the trunk of an aspen tree, photographed at night illuminated with either white light or ultraviolet-A radiation. Camera: Olympus E-M1 Mk II, Sigma 30 mm 1:1.4 DC DN MFT lens. A: White light illumination from a white LED light source, B: UV-A illumination (365 nm Convoy 2+ flashlight with VIS-blocking filter) and Tiffen Haze 2A + Firecrest UV400 filters on lens. Photographs white-balanced and edited in Capture One Pro version 16.5.
Can the flourescence be emitted at a shorter wavelength than the excitation?
It can in exceptional cases when the energy acquired by a molecule by absorbing more than one photon is emitted as a single photon. In such a case, even if there is thermal energy dissipation, and the emitted photon carries less energy than the sum of the absorbed photons, it will carry more energy than each absorbed photon individually. This happens only in exceptional cases because the re-emission is normally fast.
Photographing fluorescence
I have in Flicker some galleries of fluorescence images.
In many respects photographing fluorescence is similar to photographing in low-light. It usually requires long exposure-times and use of a tripod or other camera support. In addition we need to make sure that what we photograph is really fluorescence emitted by the object being photographed rather than contamination by excitation radiation, ambient light or even fluorescence by nearby objects.
Is it really fluorescence?
We can check if the light in the image is from the light source or not by including a “white” reflectance reference ([#fig-lichens-fluo-02]). I use a slab of white PTFE (Teflon) about 5 mm thick. It should have a matte surface and be clean, as dirt can easily fluoresce. (I use wet/dry sandpaper 600 grit under running water from time to time to clean it.) If the white slab appears darker, nearly black, in an otherwise brighter image, we can conclude that what has been recorded is fluorescence.
A: In white light.
B: In ultraviolet-A radiation.
Figure 6: Lichens on a rock. To the left of the rock the white PTFE slab can be seen. Camera: Olympus E-M1 Mk II, Sigma 30 mm 1:1.4 DC DN MFT lens. A: White light illumination from a white LED light source, B: UV-A illumination (365 nm Convoy 2+ flashlight with VIS-blocking filter) and Tiffen Haze 2A + Firecrest UV400 filters on lens. In B: a weak blue-violet shine on the reference slab was subtracted from the whole image. Photographs white-balanced and edited in Capture One Pro version 16.5.
Some equipment and suitable surroundings are needed when photographying fluorescence. The items below are discussed in more detail in later sections.
Location, outdoors: At night, far from any artificial illumination in a moon-less night.
Location, indoors: In a dark room, with no strongly fluorescent objects except the one being photographed.
A source of UV-A radiation: A flash light based on a good UV-A LED is very convenient, especially outdoors. Indoors, mains powered LED or fluorescent “black light” lamps could be also used.
Exciter filter: A high-quality VIS-blocking filter is needed in the light source.
Barrier filter: A high-quality UV-blocking filter is needed on the camera lens.
Camera: A digital mirrorless (ML) system camera is preferable, but in principle any camera whose controls allow long exposures is enough. A converted camera is needed only to photograph NIR fluorescence.
Objective: Most lenses work, but a lens with an f/2.8 or wider maximum apperture is best.
Support: Exposure times between 20 s and a few minutes are common, so the camera must be mounted so that it does not move.
Subject: Easier to photographs are several lichens that fluoresce brightly, tolerate handling and do not wilt quickly. Of course, chlorophyll fluorescence can be strong under some conditions. Blue fluorescence is common in grasses.
Location and surroundings
As fluorescence is weak compared to other light sources it can only be observed and photographed in the absence of other sources of visible light. Outdoors, normally at night and far from street lamps and other light sources, or in a darkened room indoors. The main problem indoors is the presence of fluorescent objects other than those of interest. These include paper, light-coloured and white cloths and clothes, and even some plastic objects. The reason why UV-induced blue fluorescence is pervasive indoors is that blue-fluorescing compounds are normally added to white paper and some white plastics so that they look whiter and brighter. In fact, these chemical compounds are frequently called whiteners. Many laundry powders and liquids contain whitening and/or brightening agents that remain on clothes after washing them. Less frequently, white and light-coloured paints can be formulated with fluorescent substances in addition to white pigments that reflect visible light.
Note
Post-it and similar yellow notes do not fluoresce much when exposed to UV-A radiation and can be used for labels or for taking notes.
UV-A sources
I find UV-A LEDs most useful as UV-A sources as they intrinsically emit little VIS radiation. Little is anyway in most cases too much when photographing fluorescence and they need to be combined with optical filters that effectively block visible radiation. Flash lights and other small sources of UV-A are readily available as they are used for checking authenticity of bank notes, cleanliness (urine fluoresces strongly) and searching for scorpions in houses. LEDs emitting at a wavelength of 365 nm are the most suitable (Figure 7), as radiation at this wavelength is weakly visible.
LEDs emitting at longer wavelengths are more difficult to use because we see these wavelengths better, which makes seen the fluorescence more difficult. A key difficulty is finding a good enough barrier filter at say 450 nm or 480 nm. Such filters exist, but being special can be very expensive, but required to block violet and blue light that such LEDs emit.
LEDs emitting at wavelengths shorter than 360 nm tend to have much weaker UV emission, making them cumbersome for regular use. These LEDs are also more expsnive and require more care to ensure safety.
Figure 7: Emission spectrum of an UV-A LED with nominal peak of emission at 365 nm. LED type 3W SMD LED type LZ1-10UV00 from LED Engin.
Although the emission of LEDs takes place in a relatively narrow peak, the tails can extend tens of nanometres on either side of the wavelength at the maximum emission (Figure 8). Furthermore, the peak wavelength can shift by a few nanometres as the LED temperature and/or current used to drive it changes.
Code
autoplot(leds.mspct$LedEngin_LZ1_10UV00_365nm, geom ="spct", range =c(315, 450))
Figure 8: Emission spectrum of a bare UV-A LED with nominal peak of emission at 365 nm. LED type 3W SMD LED type LZ1-10UV00 from LED Engin OSRAM. Close view of wavelength-range surounding the peak of emission.
A flashlight with a Nichia UV-A1 LED with nominal peak of emission at 365 nm, filtered with a VIS blocking filter has a much shorter right-side tail.
Figure 9: Emission spectrum of a Convoy 2+ flashlight with a 3W Nichia LED emitting at 365 nm and a ZWB2 UV-pass VIS-blocking replacement filter installed..
Black-light-blue and UV-A fluorescent tubes can also be used indoors, or where mains power is available. They usually emit over a broader range of wavelengths than individual UV LEDs, something that can sometimes be an advantage (Figure 10). Many of them emit a significant fraction of the total radiation in the visible. This makes fluorescent tubes much less suitable as blocking the unwanted VIS and NIR radiation is more difficult than with UV-A LEDs.
Figure 10: Emission spectra of black-light-blue (BLB) fluorescent tube and of a UV-A fluorescent tube.
In principle a photography flash modified to emit UV-A radiation could be used as a light source. Xenon lamps do emit over a very broad range of wavelengths, making them very difficult to filter effectively (Figure 11).
Figure 11: Emission spectra Xenon flash lamp with a glass envelope and no filter. Godox AD200 flash with a non-standard flash bulb in a bare-lamp head.
To be able to photograph fluorescence isolated from VIS and NIR radiation, without contamination from the UV-A sources, it is necessary to use UV-pass VIS-blocking filters on the light source. This barrier filter should completely block the excitation light making a high-quality UV-blocking filter necessary. Most UV-filters sold for photography, even from well-known brands, have a cut-in at too short wavelengths (330 to 380 nm). However, when using a filtered 365 nm LED as light source, the filter used on the lens has to block wavelengths up to just past 400 nm.
The best, but rather expensive, option is a Zeiss UV T* filter, second best is a significantly cheaper Firecrest UV400 filter (from Formatt-Hitech, apparently gone bankrupt), and a third option is a Tiffen Haze 2A filter (rather difficult to find outside USA). The first two are interference filters and reflect UV, rather than absorb it. The Tiffen filter, is of a rather old type and in the tradition of Kodak Wratten filters, has a thin UV-absorbing gelatine layer encased between two glass sheets. None of these three filters fluoresce significantly when exposed to UV radiation. In contrast, as discussed later in this post, most yellow, orange and red filters are made of ionic glass and when illuminated with UV radiaiotn fluoresce strongly. Thus, they can be used only behind a UV-blocking filter. Common NIR pass filters used in photography are also ionic and do also fluoresce, but less intensively.
Figure 12: Comaprison of the cut-in region of two common UV-blocking photography filters. The Zeiss UV T* filter has a very sharp change in transmittance and is extremely effective in blocking all radiation shorter than 405 nm. The Hoya UV(0) HMC filter has a much more gradual cut-in and effectively blocks wavelengths shorter than 355 nm.
When combined with the emission spectrum of the Convoy 2+ flashlight, the Zeiss filter “leaks” only \(\approx 0.03\%\) of the UV from the flash light, while the Hoya filters “leaks” \(\approx 6.5\%\) (Figure 13). This leak is more than enough to interfere with the detection of fluorescence. How much of this radiation reaches the camera sensor depends on the lens, but in any case it would be a major problem.
Figure 13: Comaprison of the cut-in UV-blocking by two different photography filters. The Zeiss UV T* filter has a very sharp change in transmittance and is extremely effective in blocking all radiation from the Convoy 2+ UVA flashlight. The Hoya UV(0) HMC filter has a much more gradual cut-in and allows through a significant part of the UVA emitted by the flashlight.
UV-A flashlights are onvenient excitation sources for use in the field (Figure 15, Figure 16). When the working distance is relatively short or even illumination is desired, a flood flashlight works better. In the case of long exposure times and use of “light painting” the more concentrated beam from the Convoy 2+ allows better targeting of the UV radiation. The Convoy 2+ is also suitable for illuminating small objects. Both flashlights have Nichia LEDs with a nominal peak of emission at 365 nm. The exact types were not specified when I bought the flashlights, but at least the Jaxman Uc1 most likely has a NVSU233B(T) LED from Nichia.
A: Convoy S2+ with a 3 W Nichia UV-A 365nm LED.
B: Jaxman Uc1 flashlight with a 6 W UV-A 365nm LED attached to an ad-hoc holder.
C: Jaxman Uc1 flashlight with a 6 W UV-A 365nm LED attached to an ad-hoc holder.
D: Jaxman Uc1 flashlight with a 6 W UV-A 365nm LED detail view of the VIS-blocking filter.
Figure 14: Photographs of two very similarly looking UV-A flashlights, that differ mainly on the LED power and optics. The Jaxman is a “flood” light with a broad and even beam and the Convoy is a “spot” light with a narrower beam that is less even. Both flashlights were bought from AliExpress sellers. The Jaxman was delivered with the VIS-blocking filter installed, and the Convoy was delivered with a clear glass that I replaced with a filter bought separately, also from an AliExpress seller.
The emission spectrum of Convoy 2+ flashlight ([Figure 15) and of the Jaxman Uc1 flashlight (Figure 16) have no stray light at longer wavelengths. Both flashlights use a single Lithium battery of type 16850, that needs to be removed for recharging.
Figure 15: Convoy 2+ flashlight with a 3W Nichia LED emitting at 365 nm and a ZWB2 UV-pass VIS-blocking replacement filter installed. This flashlight emits a narrow beam of high intensity light with a very obvious hot-spot at the centre.
Figure 16: Jaxman U1c flashlight with a 6W LED emitting at 365 nm with a built-in filter of unknown brand and type. This is a “flood” flashlight with broader and more evenly illuminated beam.
Choosing an UV-A flashlight
There are many different UV-A or black-light flashlights available. Most suitable are those emitting at 365 nm, because those emitting at 385 nm and longer wavelengths emit much more visible light (Figure 17), that will either interfere with the recording of the visible fluorescence or be blocked by the necessary UV+blue-blocking filter. The flashlight in this example emitted much less radiation and at longer wavelength.
It is also important to consider the beam uniformity and breadth. Finally, the power of most flashlights is not given by sellers as energy flux of UV-A radiation but instead as electrical power consumed. In some cases the power values given are clearly inflated, possibly based on the maximum power the LEDs withstand rather than the power at which they work in the flashlight or are simply imaginary, i.e., a hand-held flash light with a rating of 200 W if true would burn the user’s hand in no time!
Figure 17: Emission spectrum of a UV-A flashlight that was found very unsuitable in comparison to the Convoy 2+ with Nichia LED.
Convoy 2 and 2+ flashlights
The Convoy 2 and 2+ flashlight have been available from some years. They are modular, available in many different configurations and even as parts for self assembly. Flashlight using the same metal body and other parts are sold under other brands. For example, the Jaxman is very similar to my Convoy 2+ but it has, apparently, a different LED and or optics, and a fifferent LED driver.
The same body Convoy 2+ is sold with different LEDs installed, different wavelengths, different powers, and with different packages. There are two types of reflectors, clear glass, AR-coated glass, and UV-pass VIS-block windows. Drivers differ in the maximum current, that needs to be matched to the LED used, and in the user interface. The user interface can allow dimming or not, and even in some cases have a bliking option. Some drivers have a single user interface and others have multiple ones to choose from. One has to be careful with the driver and LED, as a driver that feeds a very high current to the LED will heat the flashlight. Several of the drivers have over-heating protection that decreases the power as needed to avoid overheating. This works when use is briefly, but for a longer photography session one want constant illumination.
All in all it is best to buy a ready assembled Convoy flashlight with Nichia UV LED with peak at 365 nm and an UV-pass filter. It is possible, however, to buy the parts separately, even from different suppliers and assemble a custom version with a specific LED of one’s choice. When I bought the Convoy 2+ it was not available with a pre-installed filter. Nowadays there are multiple UV-A versions, and those using LEDs from other brands than Nichia are cheaper.
Configurations ready available are with white light of different colour temperatures and powers, UV-C. UV-A, blue, green, read and infra-red. Convoy has an official store in AliExpress (“Convoy Flashlight Store”) that has this and several other flashlights.
The standard configuration uses a 18650 rechargeable Lithium battery. The body of the flashlight has a diameter of about 20 mm and is 80 mm long. To hold it I am using a holder that is sold to attach small cameras to the handlebar of a bicycle. It can be found in AliExpress under the name of “1/4 Screw Metal Cycling Bike Mount Motorcycle Handlebar Holder for GoPro 13 12 11 10 9 8 Insta360 X3 X4 DJI Osmo Action 4 5 Pro” by multiple sellers. This adds as attachment point a 1/4”-20 male thread, the standard size used for cameras. In the photograph, I have attached a male “flash cold shoe”.
Flashlight with zoom optics
In the second half of 2025 a UV-A1 flashlight model with zoom optics has become available both through AliExpress sellers, very cheapply at \(\approx 8\) €, and through a supplier of equipment for UV and IR photography in the USA at a higher price of \(\approx 20\) €. A zoom optics has a rather thick lens instead of a thin glass window or filter. To be useful in UV-A1 a lens needs to transmit wavelengths down to 340 nm well. I haven’t found any spectra or specifications for the LED or its power. The cheap and expensive flashlights look identical in the seller’s photographs. Flashlights with zoom optics have earlier been available for visible light and NIR radiation. I have one of each, and specially the white light one from LedLenser is excellent and very convenient to use.
The brand of the cheap ones is AloneFire, and reviews of another flashlight of this brand are negative, indicating badly designed electronics and cooling. The higher priced version from Kolari Vision could potentially have different electronics in the same body. By comparison, the Convoy 2+ with 3W Nichia LED costs \(\approx 35-45\) € in AliExpress and the Jaxman Uc1 with 6W Nichia LED and flood optics \(\approx 70-75\) € in AliExpress.
Exciter filter
As mentioned above, if we want to photograph fluorescence across the whole visible range, the light source used for excitation should not emit any radiation in the visible range. In fact, more generally we need to use complementary filters in the light source and camera. Most frequently we aim at a cross-over of the transmittances so that neither of the two filters transmits at 400 nm. In practice, we need a “dead-zone” as the cut-off and cut-in wavelengths of filters have tolerances of a few nanometres and there can be a gradual change in transmittance within a range of wavelengths. We look first at the filters commonly used on light sources or “exciter filters”.
Digital cameras have built-in filters on their sensors, but these filters frequently let through some UV-A radiation. Additional filters must be chosen considering the light source in use. For example, many UV-pass filters are not effective at blocking the near infra-red radiation (NIR) that some cameras can “see” (Figure 18). Fluorescent lamps emit some NIR radiation while UV-A LEDs do not emit it (cf. Figure 10, Figure 7).
Figure 18: Spectral transmittance of two band-pass UV filters with a thickness of approximately 2 mm. Both of these glass filters are not anti-reflexion coated.
Filter types ZWB1 and ZWB2 from various Chinese suppliers are cheaper than the German made Schott UG1 and UG5. Specifications are less tight for the Chinese filters, and can sometimes present optical defects. ZWB1 and ZWB2 are good filter types for use with UV-A LEDs. ZWB1 with a cut-off at a slightly shorter wavelength can be preferable. A filter thickness of 2 mm or more is necessary. These filters are usually fine if bought from reliable suppliers. Filters ready cut to size for common flashlight types are readily available through AliExpress and eBay sellers. In LED UV-A light sources it is best to use a VIS-blocking filter.
Visible light as excitation
If we are interested in red or NIR fluorescence we can use blue- or UV + blue radiation for excitation, with suitable filters and LEDs or lamps.
The Convoy 2+ flashlight is available in innumerable variations, with different LEDs, different LED drivers, and different reflectors. Even the parts can be bought as a key, and as it uses a standard sized metal core printed circuit board for a 3W LED, it would be possible to assemble one with almost any SMD 3W LED, or even 6W LED. However, instead of assembling one from parts, I bought a ready-made one with a high power blue LED.
Figure 19: Convoy 2+ flashlight with a 3W-rated OSRAM KB CSLNM1.14 LED emitting at 450 nm and a clear AR coated window installed. This flashlight emits a narrow beam of very high intensity light with a very obvious hot-spot at the centre.
As with ionic absorptive UV short-pass filters, blue short-pass filters block NIR very poorly (Figure 20). So, once again, easiest is to use LEDs as blue-light excitation sources.
Figure 20: Transmittance spectra of two types of blue ionic glass filters.
Differences in blocking efficiency among reflective dichroic filters can be huge (Figure 21). The differences in price are also huge, in the case of these two filters, 20 € compared to 190 € for filters 25 mm in diameter.
Figure 21: Transmittance spectra of two types of blue dichroic glass filters. A relatively low cost one from UQG Optics and a top-of-the line filter from Thorlabs.
Barrier filter
Barrier filters are used in front or behind the camera lens to block the excitation light completely. In this case there is an additional difficulty, yellow, orange and red glass filters themselves fluoresce when exposed to UV radiation. The solution is to use a UV-blocking filter that either reflects UV-radiation in front of the glass filters (Figure 22). These UV-blocking filters can be, as discussed above, dichroic interference filters or filters based on a different pigment than the metal ions used in solid glass filters. Kodak publications from the film era recommend using a Kodak Wratten 2A or 2B filter in front of other barrier filters. Tiffen still sells filters very similar to the Kodak ones. In these filter a coloured light-absorbing film, in most cases using gelatine as medium for the pigment, is encased between two layers of clear optical glass. These filters used to be also available as bare gelatine films. The Zeiss UV T* filter is based on interference and reflects radiation below 405 nm, it is AR coated and less affected by reflections at the transmitted wavelengths than the Tiffen. The Zeiss T* filter prevents the fluorescence of other glass filters as effectively or better than the Tiffen Haze 2A, but it is rather expensive. The Firecrest UV400 multicoated filter is nearly as good as the Zeiss, but significantly cheaper. In actual use, these three filters work well.
Figure 22: Two UV-blocking filters that do not fluoresce when exposed to UV-radiation. A Tiffen Haze 2A, equivalent to Kodak Wratten 2A and Zeiss UV T-* filter. Only the Zeiss filter has an anti-reflection coating.
In some cases we may want to photograph fluorescence only of some colours. High quality, dichroic band-pass filters would be perfect for this, but at reasonably large diameters, are very expensive. These filters are the ones used to separate fluorescence from different dyes in microscopy. If the aim is to detect all fluoresnce at wavelength longer than a target, it is possible to use a long-pass filter. If this is a ionic glass filter, mounted behind a UV-blocking filter (Figure 23).
Figure 23: Spectral transmittance of some long-pass VIS absorptive ionic glass filters. All these filters have anti-reflexion multi-coating.
With a camera modified to “see” a broader spectrum by replacement of the sensor filter, it is possible to photograph NIR fluorescence by blocking visible fluorescence with long-pass barrier filters (Figure 26). Compared to an unmodified camera leaks of NIR from the excitation radiation source become an even bigger problem. Also NIR long-pass filters require “protection” from UV to prevent their fluorescence.
Fluorescence of glass filters
Yellow, orange, red and NIR long-pass glass filters fluoresce when exposed to UV-A radiation, most strongly the yellow filters (Figure 24). Yellow filters do emit a very strong yellow fluorescence. The fluorescence of orange and specially red filters is not so intense. The fluorescence of NIR filters is mainly NIR and not visible to a normal camera or the naked eye. Using a full-spectrum converted camera and a long-pass filter on the lens, we can see that all four glass filters fluoresce in the IR.
Figure 24: Photographs of four long-pass glass filters under white visible and ultraviolet-A radiation. The background is in all photographs the same slab of white PTFE (Teflon), but in B–D the filters used on the camera lens blocked the UV-A radiation used to excite fluorescence. Clockwise from the top left corner: Hoya Red 25A (603 nm), Hoya R72 (716 nm),Heliopan Orange 22 (568 nm) and Heliopan Yellow 5 (466 nm), all AR coated. Light sources: Sunwayfoto FL96 at 5500K LED fill light and Jaxman Uc1 flashlight. Cameras A, B: Olympus OM-1; C, D: Olympus E-M1 full-spectrum converted. For visible fluorescence barrier filter: Firecrest UV400 + Tiffen 2A, and for NIR fluorescence Firecrest UV400 + C:Zomei IR680 or D:Haida IR720 (726 nm) + Tiffen 2A. Lens: M.Zuiko 25 mm 1:1.2 Pro. Photographs white-balanced and edited in Capture One Pro version 16.5.
The spectra of the fluorescence emission have one or two peaks depending on the filter, and these peaks are at longer wavelengths than those of the radiation used for excitation (Figure 25).
Figure 25: Fluorescence spectra for selected photography filters under excitation from 900 umol m-2 s-1 UV-A1 radiation with peak at 366 nm. A Convoy S2+ flashlight with 3W Nichia LED and a VIS-blocking filter was used as light source (spectrum in Figure 15). The text header of each panel indicates supplier, filter type and filter thickness.
To control fluorescence from these filters an additional non-fluorescent UV-reflecting or absorbing filter must be stacked in front of them. Examples of effective UV-blocking filters are Zeiss UV*, Firecrest UV400 and Tiffen Haze 2A. Fluorescence can be also excited by blue light, so UV-blocking filters do not always prevent fluorescence from yellow, orange and red glass filters.
Figure 26: Spectral transmittance of some long-pass NIR filters. These are all glass filters, of which the Hoya R72 is anti-reflexion coated.
In principle band-pass filters can also be used, although this is a specialized situation. Of band-pass filters, the only sold for photography ready mounted on rings are the UV-IR cut filters. The actual wavelength boundaries vary rather broadly among brands and types (Figure 27). In AliExpress and eBay sometimes unusual ones are offered.
Figure 27: Transmittance spectra of an assortment of UVIR-cut filters sold for photography and video.
Filters can be stacked, and this is frequently necessary with UV-pass filters that also transmit some NIR radiation.
A huge selection of band-pass filters mounted on suitable rings are available from suppliers of equipment for industrial machine vision. For example, MIDOPT (distributed in Europe by Stemmer Imaging), is a well known supplier, but such filters tend to be more expensive than those sold for photography. Another reliable supplier with a large variety of filters is Thorlabs, including filters specifically made for fluorescence imaging in microscopy, but mainly in smaller sizes. Square filters in a 2” \(\times\) 2” or 50 mm \(\times\) 50 mm size are common in the catalogues of optics suppliers like Hebo glass, UQG Optics, and some local optics shops, e.g., Teknofokus in Finland.
Cameras and camera settings
Normal cameras can be used for photographing the visible fluorescence. They are preferable to full-spectrum converted cameras as the built-in sensor filter help block unwanted UV and IR radiation. To photograph NIR fluorescence, an IR or full-spectrum-converted camera is needed.
A key requirement is that the camera supports long exposure times, i.e., that it has at least a Bulb setting for shutter speed as exposures may extend to minutes. Cameras that can display the live image as is “builds” during a long exposure are extremely convenient. In Olympus/OM-System cameras this mode is called and is what I always use for UVIVF photography in the field at night. In the case of close-up and macro photography it is possible to use exposure times that are shorter and to illuminate the whole view evenly.
A lens with a large aperture helps, but is not a must as long as subjects do not move. Any good lens is suitable as the UV radiation is used to excite the fluorescence and we block it with a filter before it enters the lens. Thus, it does not require the especial lenses and modified cameras that are required for reflected UV photography.
White balancing fluorescence photographs
Objectively white balancing a photograph of fluorescence is nearly impossible. After trying other approaches I have mostly settled into editing the colour in these photographs to “look right”, i.e., matching my recollection of how the subject looked like when I photographed it. However, in some cases I use just a daylight balance and alternatively edit the colour to remove intense casts so as to increase the apparent colour range.
Close-up photography of fluorescence
Combining the close-up or macro-photography with ultraviolet-induced visible fluorescence in the field can be difficult as exposure times are long and focus stacking is frequently needed for subjects with relief. A few examples from last night can be seen in Figure 28.
A: In visible light.
B: UV-A-induced visible fluorescence.
C: In visible light..
D: UV-A-induced visible fluorescence.
E: In visible light..
F: UV-A-induced visible fluorescence.
G: In visible light..
H: UV-A-induced visible fluorescence.
I: In visible light..
J: UV-A-induced visible fluorescence.
K: In visible light..
L: UV-A-induced visible fluorescence.
Figure 28: Photographs of four dead tree branches, covered with lichens and mosses. The background is in all photographs the same black velvet cloth. The edge of a slab of white PTFE (Teflon) is visible on the top-right corner of the photographs. In, B, D, F, H, and J the filters used on the camera lens blocked the UV-A radiation used to excite fluorescence. Light sources: Sunwayfoto FL96 at 5500K LED fill light and Jaxman Uc1 flashlight. Camera: Olympus OM-1. For visible fluorescence barrier filter: Firecrest UV400 + Tiffen 2A. Lens: A–J M.Zuiko 25 mm 1:1.2 Pro, K-L M.Zuiko 90 mm 1:3.5 Macro Pro. Photographs white-balanced and edited in Capture One Pro version 16.5. Focus-stacks of 20 or 80 photographs were merged in Helicon Focus 8.2.18 Pro to increase the depth of field.
With suitable filters and lenses we can even attempt macro-photography of UV-A-induced chlorophyll fluorescence with a modified mirrorless camera (Figure 29).
A: In visible light.
B: UV-A-induced near-infrared flousrescence.
Figure 29: Moss on a dead branch. Illuminated with A: white light or B: UV-A radiation. Only difference with images in Figure 28 in the approach was adding a long-pass filter with a cut-in at 650 nm (HB650). Using a filter with cut-in at 695 nm (Heliopan RG695) resulted in a similar image but required a longer exposure time. Camera: full-spectrum converted Olympus E-M1 camera. See legend to Figure 28 for other details.
Conclusion
Photographing UV-A-induced visible fluorescence is interesting and less demanding in equipment and technique than photography in reflected UV radiation. An unmodified digital camera can be used and long pass filters are in general cheaper and more common than UV short-pass filters that block visible and infrared radiation effectively enough. UV-A as excitation is convenient and easy to work with, blue light could be used instead if we are interested in fluorescence at longer wavelengths or for safety reasons in teaching situations. Using UV-B radiation to induce UV-A and violet fluorescence is in principle possible but more demanding in practice. Photographing visible-light induced near infrared fluorescence is more demanding in equipment, but opens interesting possibilities. Could we with comparatively cheap equipment use photography to qualitatively assess changes epidermal UV transmittance under controlled conditions?
Other sources of information and help
The UltravioletPhotography discussion group is a very friendly and knowledgeable group of people from around the world with an interest in UV, NIR, and UV-induced fluorescence.
The web sites of Adrian Davies, author of three of the books listed below, include galleries of photographs, including some with photographs of UV-A induced fluorescence. See Imaging the Invisible and Adrian Davies imaging.
Further reading
Blacklock, Craig, and Nadine Blacklock. 1990. Photographing Wildflowers: Techniques for the Advanced Amateur and Professional. Stillwater, Minnesota: Voyageur Press.
Davies, Adrian. 2017. Digital Ultraviolet and Infrared Photography. Focal Press.
———. 2020. Photographing the Unseen World: Art and Techniques. Ramsbury, Wiltshire: Crowood Press.
Gauslaa, Y., and E. M. Ustvedt. 2003. “Is Parietin a UV-B or a Blue-Light Screening Pigment in the Lichen _Xanthoria Parietina_?”Photochemical & Photobiological Sciences 2 (4): 424–32. https://doi.org/10.1039/b212532c.
Lindfors, Anders V., and Lasse Ylianttila. 2016. “Visualizing Rayleigh Scattering Through UV Photography.”Bulletin of the American Meteorological Society 97 (9): 1561–64. https://doi.org/10.1175/BAMS-D-14-00260.1.
---title: "UV-A-Induced VIS Fluorescence---Methods"subtitle: "UV-A sources, filters, cameras and other equipment"author: "Pedro J. Aphalo"date: 2022-10-15date-modified: 2025-11-22toc: truecategories: [equipment, illumination, filters]keywords: [LED light, ultraviolet]bibliography: uvaivf-methods.bibnocite: | @*format: html: code-fold: true code-tools: truelightbox: autoimage: images/lichens-UVIVF-01.jpglicense: "CC BY-SA"draft: falseeditor: markdown: wrap: 72abstract: | This page contains notes about photography of ultraviolet-A-radiation induced visible fluorescence. I discuss methods in detail and provide examples of higher plants, lichens and mosses. I describe the ultraviolet-A sources, UV-pass and UV-blocking filters, lenses and cameras that I use, as well as camera settings I use.---{{< include /_includes/uva-warning.qmd >}}::: callout-tip# Viewing photographsClicking on photographs opens a gallery view with them displayed at alarger size and at higher resolution. This view also allows navigationamong multiple photographs from a given Figure.:::{{< include /_includes/folded-code-tip.qmd >}}```{r, message=FALSE}library(photobiology)library(photobiologyFilters)library(photobiologyLamps)library(photobiologyLEDs)library(photobiologyWavebands)library(photobiologyPlants)library(ggspectra)library(patchwork)library(knitr)my.fig.heigth <-4.5my.fig.width <-7knitr::opts_chunk$set(fig.width = my.fig.width, fig.height = my.fig.heigth, dev ="svg", out.width ="975%")theme_set(theme_bw(12) +theme(legend.position ="top"))photon_as_default()```## FluorescenceFluorescence occurs when a molecule absorbs radiation and re-emits theabsorbed energy also as radiation. The usual case is the capture of asingle photon, dissipation of a (small) fraction of the energy as heatand the emission of the rest of the energy as a new photon. As theemitted photon carries less energy than the absorbed one, fluorescencetakes place at a longer wavelength than the absorbed radiation.Both UV and visible light can induce the fluorescence of chlorophyll inplants. Strong fluorescence is emitted at wavelengths near 685 nm and740 nm, by chlorophyll-_a_ and chlorophyll_b_. Phenolic acids fluorescein the blue and green regions. Pigments in lichens, some mushrooms and even insects (as well as scorpions) fluoresce in a variety of bright purecolours.We do not see the fluorescence in normal daylight as the highsensitivity of human vision with a peak in green prevent it. With acamera modified to increase sensitivity to infra-red radiation and afilter that blocks wavelengths shorter than 680--700 nm the fluorescencecan be imaged. A flash of strong light is most effective and is used inthe study of plant's photosynthesis ([@fig-faba-chl-fluorescence]).::: {#fig-faba-chl-fluorescence}{fig-alt="Two young fava bean plants glowing orange-red on a dark background."}NIR chlorophyll fluorescence induced by a flash of white light in plantsof fava beans. Camera: full-spectrum converted Olympus E-M1, Sigma 30 mm1:2.8 DC DN MFT lens, Heliopan RG680 filter on lens. Strong white lightillumination from a device used to study photosynthesis of plants.:::The fluorescence spectrum can be measured, with some difficulty withan array spectrometer and a UV-A flashlight resulting in very noisy data. ```{r, warning=FALSE, message=FALSE}#| label: fig-lettuce-leaf-fluo-spct#| fig-cap: Whole-leaf fluorescence spectrum from a lettuce leaf illuminated with UV-A1 radiation at 365 nm from a flashlight and measured on the opposite side of the leaf with a Maya 2000 Pro array spectrometer with a cosine diffuser. (Heavily smoothed.)load("data/UVA365nm_source_003.spct.Rda")autoplot(clean(smooth_spct(UVA365nm_source_003.spct, method ="supsmu", na.rm =TRUE)),geom ="spct",span =55,range =c(400, 800),label.qty ="contribution.pc")```Of course it can be also measured withmore sophisticated equipment to obtain a clean fluorescence spectrum, in this case for a wheat leaf, measuredunder 355 nm excitation ([@fig-wheat-leaf-fluo-spct]).```{r, warning=FALSE, message=FALSE}#| label: fig-wheat-leaf-fluo-spct#| fig-cap: Whole-leaf fluorescence spectrum from a wheat leaf illuminated with UV-A1 radiation at 355 nm.autoplot(leaf_fluorescence.mspct, span =11, geom ="spct")```The quantum yield of fluorescence depends on the molecules involved. Inmost cases the fraction of the incident radiation that is emitted asfluorescence is small, frequently not even detectable. When weilluminate a plant or object with ultraviolet-A (UV-A) radiation, onlysome of its components, if any, fluoresce strongly, emitting visiblelight (VIS) and NIR radiation of specific wavelengths.To induce fluorescence across the whole visible spectrum we haveto use radiation of shorter wavelengths, such as UV-A radiation. Oureyes' low sensitivity to UV-A radiation and easily available eyeglassesand goggles that block all UV radiation help in allowing us to see thefluorescence without it being overwhelmed by the radiation used asexcitation.Different lichens contain pigments that fluoresce brightly when exposedto UV-A radiation ([@fig-lichens-fluo-01]). Colours of the fluorescencevary between yellow, red, occasionally blue and rarely green.::: {#fig-lichens-fluo-01 layout-ncol="2"}{fig-alt="Trunk of an aspen tree in white light with yellow and grey lichens."group="lichens01"}{fig-alt="Trunk of an aspen tree in UV-A with the same lichens glowing red and yellow."group="lichens01"}Lichens on the trunk of an aspen tree, photographed at night illuminatedwith either white light or ultraviolet-A radiation. Camera: Olympus E-M1Mk II, Sigma 30 mm 1:1.4 DC DN MFT lens. A: White light illuminationfrom a white LED light source, B: UV-A illumination (365 nm Convoy 2+flashlight with VIS-blocking filter) and Tiffen Haze 2A + FirecrestUV400 filters on lens. Photographs white-balanced and edited in CaptureOne Pro version 16.5.:::::: callout-note# Can the flourescence be emitted at a shorter wavelength than the excitation?It can in exceptional cases when the energy acquired by a molecule byabsorbing more than one photon is emitted as a single photon. In such acase, even if there is thermal energy dissipation, and the emitted photoncarries less energy than the sum of the absorbed photons, it will carrymore energy than each absorbed photon individually. *This happens onlyin exceptional cases because the re-emission is normally fast.*:::## Photographing fluorescenceI have in Flicker some galleries of fluorescence images.[Lichens photographed at the Jardin botanique du col du Lautaret, French Alps, 2019](https://flic.kr/s/aHsmDXGTvR)[Various photographs taken indoors, at home, Helsinki, Finland](https://flic.kr/s/aHsmUGu7MG)[Various photographs taken outdoors at night, Loppi, Finland, 2019-2020](https://flic.kr/s/aHBqjAuFCM)In many respects photographing fluorescence is similar to photographingin low-light. It usually requires long exposure-times and use of atripod or other camera support. In addition we need to make sure thatwhat we photograph is really fluorescence emitted by the object beingphotographed rather than contamination by excitation radiation, ambientlight or even fluorescence by nearby objects.:::: callout-tip# Is it really fluorescence?We can check if the light in the image is from the light source or notby including a "white" reflectance reference (\[#fig-lichens-fluo-02\]).I use a slab of white PTFE (Teflon) about 5 mm thick. It should have amatte surface and be clean, as dirt can easily fluoresce. (I use wet/drysandpaper 600 grit under running water from time to time to clean it.)If the white slab appears darker, nearly black, in an otherwise brighterimage, we can conclude that what has been recorded is fluorescence.::: {#fig-lichens-fluo-02 layout-ncol="2"}{fig-alt="Rock in white light with grey lichens and the reference slab looking white."group="rock01"}{fig-alt="Rock in ultraviolet-A radiation with lichen glowing yellow and the reference slab looking almost black."group="rock01"}Lichens on a rock. To the left of the rock the white PTFE slab can beseen. Camera: Olympus E-M1 Mk II, Sigma 30 mm 1:1.4 DC DN MFT lens. A:White light illumination from a white LED light source, B: UV-Aillumination (365 nm Convoy 2+ flashlight with VIS-blocking filter) andTiffen Haze 2A + Firecrest UV400 filters on lens. In B: a weakblue-violet shine on the reference slab was subtracted from the wholeimage. Photographs white-balanced and edited in Capture One Pro version16.5.:::::::Some equipment and suitable surroundings are needed when photographing fluorescence. The items below are discussed in more detail in latersections.- Location, outdoors: At night, far from any artificial illumination in a moon-less night. - Location, indoors: In a dark room, with no strongly fluorescent objects except the one being photographed.- A source of UV-A radiation: A flash light based on a good UV-A LED is very convenient, especially outdoors. Indoors, mains powered LED or fluorescent "black light" lamps could be also used.- Exciter filter: A high-quality VIS-blocking filter is needed in the light source.- Barrier filter: A high-quality UV-blocking filter is needed on the camera lens.- Camera: A digital mirrorless (ML) system camera is preferable, but in principle any camera whose controls allow long exposures is enough. A converted camera is needed only to photograph NIR fluorescence.- Objective: Most lenses work, but a lens with an f/2.8 or wider maximum aperture is best.- Support: Exposure times between 20 s and a few minutes are common, so the camera must be mounted so that it does not move.- Subject: Easier to photographs are several lichens that fluoresce brightly, tolerate handling and do not wilt quickly. Of course, chlorophyll fluorescence can be strong under some conditions. Blue fluorescence is common in grasses.### Location and surroundingsAs fluorescence is weak compared to other light sources it can only beobserved and photographed in the absence of other sources of visiblelight. Outdoors, normally at night and far from street lamps and otherlight sources, or in a darkened room indoors. The main problem indoorsis the presence of fluorescent objects other than those of interest.These include paper, light-coloured and white cloths and clothes, and even some plastic objects. Thereason why UV-induced blue fluorescence is pervasive indoors is thatblue-fluorescing compounds are normally added to white paper and somewhite plastics so that they look whiter and brighter. In fact, thesechemical compounds are frequently called whiteners. Many laundry powdersand liquids contain whitening and/or brightening agents that remain onclothes after washing them. Less frequently, white and light-colouredpaints can be formulated with fluorescent substances in addition towhite pigments that reflect visible light.::: callout-notePost-it and similar yellow notes do not fluoresce much when exposed toUV-A radiation and can be used for labels or for taking notes.:::### UV-A sourcesI find UV-A LEDs most useful as UV-A sources as they intrinsically emitlittle VIS radiation. Little is anyway in most cases too much whenphotographing fluorescence and they need to be combined with opticalfilters that effectively block visible radiation. Flash lights and othersmall sources of UV-A are readily available as they are used forchecking authenticity of bank notes, cleanliness (urine fluorescesstrongly) and searching for scorpions in houses. LEDs emitting at a wavelength of 365 nm are the most suitable([@fig-LedEngin365nm]), as radiation at this wavelength is weakly visible.LEDs emitting at longer wavelengths are moredifficult to use because we see these wavelengths better, which makesseen the fluorescence more difficult. A key difficulty is finding a goodenough barrier filter at say 450 nm or 480 nm. Such filters exist, butbeing special can be very expensive, but required to block violet and blue light that such LEDs emit. LEDs emitting atwavelengths shorter than 360 nm tend to have much weaker UV emission,making them cumbersome for regular use. These LEDs are also more expensiveand require more care to ensure safety.```{r}#| label: fig-LedEngin365nm#| fig-cap: Emission spectrum of an UV-A LED with nominal peak of emission at 365 nm. LED type 3W SMD LED type LZ1-10UV00 from LED Engin.autoplot(leds.mspct$LedEngin_LZ1_10UV00_365nm, geom ="spct")```Although the emission of LEDs takes place in a relatively narrow peak, the tailscan extend tens of nanometres on either side of the wavelengthat the maximum emission ([@fig-LedEngin365nm-zoom]). Furthermore, the peakwavelength can shift by a few nanometres as the LED temperature and/orcurrent used to drive it changes.```{r}#| label: fig-LedEngin365nm-zoom#| fig-cap: Emission spectrum of a bare UV-A LED with nominal peak of emission at 365 nm. LED type 3W SMD LED type LZ1-10UV00 from LED Engin OSRAM. Close view of wavelength-range surrounding the peak of emission.autoplot(leds.mspct$LedEngin_LZ1_10UV00_365nm, geom ="spct", range =c(315, 450))```A flashlight with a Nichia UV-A1 LED with nominal peak of emission at 365 nm, filtered with a VIS blocking filter has a much shorter right-side tail.```{r}#| label: fig-Convoy2p#| fig-cap: Emission spectrum of a Convoy 2+ flashlight with a 3W Nichia LED emitting at 365 nm and a ZWB2 UV-pass VIS-blocking replacement filter installed..autoplot(smooth_spct(normalise(lamps.mspct[c("Convoy.S2plus.LED.UVA.flashlight")], norm ="undo"), method ="supsmu", strength =0.001), geom ="spct",range =c(315, 450))```Black-light-blue and UV-A fluorescent tubes can also be used indoors, orwhere mains power is available. They usually emit over a broader rangeof wavelengths than individual UV LEDs, something that can sometimes bean advantage ([@fig-Qpanel340-lamp]). Many of them emit a significantfraction of the total radiation in the visible. This makes fluorescenttubes much less suitable as blocking the unwanted VIS and NIR radiation ismore difficult than with UV-A LEDs.```{r, warning=FALSE, fig.height=my.fig.heigth * 1.8}#| label: fig-Qpanel340-lamp#| fig-cap: Emission spectra of black-light-blue (BLB) fluorescent tube and of a UV-A fluorescent tube.autoplot(lamps.mspct[c("QPanel.FT.UVB340.40W", "Eiko.F36T8.BLB")], geom ="spct", idfactor ="Lamp", facets =1, span =101)```In principle a photography flash modified to emit UV-A radiation could be used as alight source. Xenon lamps do emit over a very broad range ofwavelengths, making them very difficult to filter effectively ([@fig-flash-spct]).```{r}#| label: fig-flash-spct#| fig-cap: Emission spectra Xenon flash lamp with a glass envelope and no filter. Godox AD200 flash with a non-standard flash bulb in a bare-lamp head.autoplot(lamps.mspct$Godox.XeF.AD200.H200.flash, geom ="spct", span =51)```To be able to photograph fluorescence isolated from VIS and NIR radiation,without contamination from the UV-A sources, it is necessary to use UV-passVIS-blocking filters on the light source. This barrier filter should completelyblock the excitation light making a high-quality UV-blocking filter necessary.Most UV-filters sold for photography, even from well-known brands, have a cut-inat too short wavelengths (330 to 380 nm). However, when using a filtered 365 nm LED as light source, the filter used on the lens has to block wavelengths upto just past 400 nm. The best, but rather expensive, option is a Zeiss UV T* filter, second best is a significantly cheaper Firecrest UV400 filter (from Formatt-Hitech, apparently gone bankrupt), and a third option is a Tiffen Haze 2A filter (rather difficult to find outside USA). The first two are interferencefilters and reflect UV, rather than absorb it. The Tiffen filter, is of a ratherold type and in the tradition of Kodak Wratten filters, has a thin UV-absorbinggelatine layer encased between two glass sheets. None of these three filtersfluoresce significantly when exposed to UV radiation. In contrast, as discussed later in this post, most yellow, orange and red filters are made of ionic glass and when illuminated with UV radiation fluoresce strongly. Thus, they can be used only behind a UV-blocking filter. Common NIR pass filters used in photography are also ionic and do also fluoresce, but less intensively.```{r}#| label: fig-uv-filters-compared#| fig-cap: Comparison of the cut-in region of two common UV-blocking photography filters. The Zeiss UV T* filter has a very sharp change in transmittance and is extremely effective in blocking all radiation shorter than 405 nm. The Hoya UV(0) HMC filter has a much more gradual cut-in and effectively blocks wavelengths shorter than 355 nm.autoplot(filters.mspct[c("Zeiss_UV_Tstar_2.0mm_52mm","Hoya_UV0_HMC_2.0mm_52mm")],range =c(315, 450)) +theme(legend.position ="top") ```When combined with the emission spectrum of the Convoy 2+ flashlight, the Zeissfilter "leaks" only $\approx 0.03\%$ of the UV from the flash light, while theHoya filters "leaks" $\approx 6.5\%$ ([@fig-uv-filters-compared-with-convoy]).This leak is more than enough to interfere with the detection of fluorescence.How much of this radiation reaches the camera sensor depends on the lens, but inany case it would be a major problem.```{r}#| label: fig-uv-filters-compared-with-convoy#| fig-cap: Comparison of the cut-in UV-blocking by two different photography filters. The Zeiss UV T* filter has a very sharp change in transmittance and is extremely effective in blocking all radiation from the Convoy 2+ UVA flashlight. The Hoya UV(0) HMC filter has a much more gradual cut-in and allows through a significant part of the UVA emitted by the flashlight.(autoplot(smooth_spct(normalise(lamps.mspct[["Convoy.S2plus.LED.UVA.flashlight"]], norm ="undo") * filters.mspct[["Zeiss_UV_Tstar_2.0mm_52mm"]], method ="supsmu", strength =0.5),range =c(315, 450), ylim =c(NA, 6.2), annotations =c("-", "peaks")) +ggtitle("Zeiss UV T* filter") +theme(legend.position ="top")) |(autoplot(smooth_spct(normalise(lamps.mspct[["Convoy.S2plus.LED.UVA.flashlight"]], norm ="undo") * filters.mspct[["Hoya_UV0_HMC_2.0mm_52mm"]], method ="supsmu", strength =0.5),range =c(315, 450), ylim =c(NA, 6.2), annotations =c("-", "peaks")) +ggtitle("Hoya UV(0) filter") +theme(legend.position ="top")) +plot_layout(axis_titles ="collect")```UV-A flashlights are convenient excitation sources for use inthe field ([@fig-convoy-spct; @fig-jaxman-spct]). When theworking distance is relatively short or even illumination is desired, a floodflashlight works better. In the case of long exposure times and use of "lightpainting" the more concentrated beam from the Convoy 2+ allows better targetingof the UV radiation. The Convoy 2+ is also suitable for illuminating small objects. Both flashlights have Nichia LEDs with a nominal peak of emission at 365 nm. The exact types were not specified when I bought the flashlights, but at least the Jaxman Uc1 most likely has a NVSU233B(T) LED from Nichia.::: {#fig-flashlight layout-ncol="2"}{fig-alt="View of a small black anodized aluminium flashlight."group="flashlights"}{fig-alt="View of a small black anodized aluminium flashlight with a support attached."group="flashlights"}{fig-alt="View of a small black anodized aluminium flashlight with a support attached."group="flashlights"}{fig-alt="Detail view of the front end of a small black anodized aluminium flashlight showing the black-looking UV-pass and VIS-blocking filter." group="flashlights"}Photographs of two very similarly looking UV-A flashlights, that differmainly on the LED power and optics. The Jaxman is a "flood" light with abroad and even beam and the Convoy is a "spot" light with a narrowerbeam that is less even. Both flashlights were bought from AliExpresssellers. The Jaxman was delivered with the VIS-blocking filterinstalled, and the Convoy was delivered with a clear glass that Ireplaced with a filter bought separately, also from an AliExpressseller.:::The emission spectrum of Convoy 2+ flashlight ([@fig-convoy-spct) and of the Jaxman Uc1 flashlight ([@fig-jaxman-spct]) have no stray light at longer wavelengths. Both flashlights use a single Lithium battery of type 16850, thatneeds to be removed for recharging. ```{r}#| label: fig-convoy-spct#| fig-cap: Convoy 2+ flashlight with a 3W Nichia LED emitting at 365 nm and a ZWB2 UV-pass VIS-blocking replacement filter installed. This flashlight emits a narrow beam of high intensity light with a very obvious hot-spot at the centre.autoplot(lamps.mspct$Convoy.S2plus.LED.UVA.flashlight, geom ="spct")``````{r}#| label: fig-jaxman-spct#| fig-cap: Jaxman U1c flashlight with a 6W LED emitting at 365 nm with a built-in filter of unknown brand and type. This is a "flood" flashlight with broader and more evenly illuminated beam. autoplot(lamps.mspct$Jaxman.U1c.LED.UVA.flood.flashlight, geom ="spct")```::: callout-attention# Choosing an UV-A flashlightThere are many different UV-A or black-light flashlights available. Mostsuitable are those emitting at 365 nm, because those emitting at 385 nmand longer wavelengths emit much more visible light ([@fig-generic-uva-flashlight-spct]), that will eitherinterfere with the recording of the visible fluorescence or be blockedby the necessary UV+blue-blocking filter. The flashlight in this exampleemitted much less radiation and at longer wavelength. It is also important to consider the beam uniformity and breadth. Finally, the power of most flashlights is notgiven by sellers as energy flux of UV-A radiation but instead aselectrical power consumed. *In some cases the power values given areclearly inflated, possibly based on the maximum power the LEDs withstandrather than the power at which they work in the flashlight or are simplyimaginary, i.e., a hand-held flash light with a rating of 200 W if truewould burn the user's hand in no time!*```{r}#| label: fig-generic-uva-flashlight-spct#| fig-cap: Emission spectrum of a UV-A flashlight that was found very unsuitable in comparison to the Convoy 2+ with Nichia LED. autoplot(lamps.mspct$Generic.LED.UVA.flashlight, geom ="spct")```:::::: callout-tip# Convoy 2 and 2+ flashlightsThe Convoy 2 and 2+ flashlight have been available from some years. They aremodular, available in many different configurations and even as parts forself assembly. Flashlights using the same metal body and other parts are soldunder other brands. For example, the Jaxman is very similar to my Convoy 2+but it has, apparently, a different LED and or optics, and a different LEDdriver.The same body Convoy 2+ is sold with different LEDs installed, different wavelengths, different powers, and with different packages. There are twotypes of reflectors, clear glass, AR-coated glass, and UV-pass VIS-blockwindows. Drivers differ in the maximum current, that needs to be matched tothe LED used, and in the user interface. The user interface can allowdimming or not, and even in some cases have a blinking option. Some drivershave a single user interface and others have multiple ones to choose from.One has to be careful with the driver and LED, as a driver that feeds a veryhigh current to the LED will heat the flashlight. Several of the drivers haveover-heating protection that decreases the power as needed to avoid overheating.This works when use is briefly, but for a longer photography session one wantconstant illumination.All in all it is best to buy a ready assembled Convoy flashlight with Nichia UV LED with peak at 365 nm and an UV-pass filter. It is possible, however, to buythe parts separately, even from different suppliers and assemble a customversion with a specific LED of one's choice. When I bought the Convoy 2+ itwas not available with a pre-installed filter. Nowadays there are multipleUV-A versions, and those using LEDs from other brands than Nichia are cheaper.Configurations ready available are with white light of different colourtemperatures and powers, UV-C. UV-A, blue, green, read and infra-red. Convoyhas an official store in AliExpress ("Convoy Flashlight Store") that has thisand several other flashlights.The standard configuration uses a 18650 rechargeable Lithium battery.The body of the flashlight has a diameter of about 20 mm and is 80 mm long.To hold it I am using a holder that is sold to attach small cameras to thehandlebar of a bicycle. It can be found in AliExpress under the name of"1/4 Screw Metal Cycling Bike Mount Motorcycle Handlebar Holder for GoPro 13 12 11 10 9 8 Insta360 X3 X4 DJI Osmo Action 4 5 Pro" by multiple sellers. Thisadds as attachment point a 1/4"-20 male thread, the standard size used forcameras. In the photograph, I have attached a male "flash cold shoe".:::::: callout-tip# Flashlight with zoom opticsIn the second half of 2025 a UV-A1 flashlight model with zoom optics has becomeavailable both through AliExpress sellers, very cheaply at $\approx 8$ €, andthrough a supplier of equipment for UV and IR photography in the USA at a higherprice of $\approx 20$ €. A zoom optics has a rather thick lens instead of a thinglass window or filter. To be useful in UV-A1 a lens needs to transmitwavelengths down to 340 nm well. I haven't found any spectra or specificationsfor the LED or its power. The cheap and expensive flashlights look identical inthe seller's photographs. Flashlights with zoom optics have earlier beenavailable for visible light and NIR radiation. I have one of each, and speciallythe white light one from LedLenser is excellent and very convenient to use.The brand of the cheap ones is AloneFire, and reviews of another flashlight ofthis brand are negative, indicating badly designed electronics and cooling. Thehigher priced version from [KolariVision](https://kolarivision.com/product/kolari-ultraviolet-flashlight-uv/)could potentially have different electronics in the same body. By comparison,the Convoy 2+ with 3W Nichia LED costs $\approx 35-45$ € in AliExpress and theJaxman Uc1 with 6W Nichia LED and flood optics $\approx 70-75$ € in AliExpress.:::### Exciter filterAs mentioned above, if we want to photograph fluorescence across thewhole visible range, the light source used for excitation should notemit any radiation in the visible range. In fact, more generally we needto use complementary filters in the light source and camera. Mostfrequently we aim at a cross-over of the transmittances so that neitherof the two filters transmits at 400 nm. In practice, we need a"dead-zone" as the cut-off and cut-in wavelengths of filters havetolerances of a few nanometres and there can be a gradual change intransmittance within a range of wavelengths. We look first at thefilters commonly used on light sources or "exciter filters".Digital cameras have built-in filters on their sensors, but thesefilters frequently let through some UV-A radiation. Additional filters must bechosen considering the light source in use. For example, many UV-passfilters are not effective at blocking the near infra-red radiation (NIR)that some cameras can "see" ([@fig-filters-UV-pass]). Fluorescent lampsemit some NIR radiation while UV-A LEDs do not emit it (cf. [@fig-Qpanel340-lamp; @fig-LedEngin365nm]).```{r}#| label: fig-filters-UV-pass#| fig-cap: Spectral transmittance of two band-pass UV filters with a thickness of approximately 2 mm. Both of these glass filters are not anti-reflection coated.autoplot(filters.mspct[c("Tangsinuo_ZWB1_2.1mm_52mm", "Tangsinuo_ZWB2_2.0mm_52mm")], geom ="spct", idfactor ="Filter", span =51)```Filter types ZWB1 and ZWB2 from various Chinese suppliers are cheaperthan the German made Schott UG1 and UG5. Specifications are less tightfor the Chinese filters, and can sometimes present optical defects. ZWB1and ZWB2 are good filter types for use with UV-A LEDs. ZWB1 with a cut-off at aslightly shorter wavelength can be preferable. A filter thickness of 2 mm ormore is necessary. These filters are usually fineif bought from reliable suppliers. Filters ready cut to size for commonflashlight types are readily available through AliExpress and eBaysellers. In LED UV-A light sources it is best to use a VIS-blocking filter.::: callout-tip# Visible light as excitationIf we are interested in red or NIR fluorescence we can use blue- or UV +blue radiation for excitation, with suitable filters and LEDs or lamps.The Convoy 2+ flashlight is available in innumerable variations, withdifferent LEDs, different LED drivers, and different reflectors. Even theparts can be bought as a key, and as it uses a standard sized metal coreprinted circuit board for a 3W LED, it would be possible to assemble onewith almost any SMD 3W LED, or even 6W LED. However, instead of assemblingone from parts, I bought a ready-made one with a high power blue LED.```{r}#| label: fig-convoy-blue-spct#| fig-cap: Convoy 2+ flashlight with a 3W-rated OSRAM KB CSLNM1.14 LED emitting at 450 nm and a clear AR coated window installed. This flashlight emits a narrow beam of very high intensity light with a very obvious hot-spot at the centre.autoplot(lamps.mspct$Convoy.S2plus.LED.blue.flashlight, geom ="spct")```As with ionic absorptive UV short-pass filters, blue short-pass filters block NIRvery poorly ([@fig-blue-filters]). So, once again, easiest is to use LEDs as blue-light excitation sources.```{r}#| label: fig-blue-filters#| fig-cap: Transmittance spectra of two types of blue ionic glass filters.autoplot(filters.mspct[c("Tangsinuo_ZB1_2.0mm_52mm", "Tangsinuo_ZB2_2.0mm_52mm")],geom ="spct", idfactor ="Filter", annotations =c("+", "wls"),span =101)```Differences in blocking efficiency among reflective dichroic filters can be huge ([@fig-blue-dichroic-filters]). The differences in price are also huge, in the case of thesetwo filters, 20 € compared to 190 € for filters 25 mm in diameter.```{r}#| label: fig-blue-dichroic-filters#| fig-cap: Transmittance spectra of two types of blue dichroic glass filters. A relatively low cost one from UQG Optics and a top-of-the line filter from Thorlabs.autoplot(filters.mspct[c("Thorlabs_FBH450_40", "UQG_Blue_dichroic_CDB")],geom ="spct", idfactor ="Filter", range =c(300, 1050),annotations =c("+", "wls"),span =101)```:::## Barrier filterBarrier filters are used in front or behind the camera lens to block theexcitation light completely. In this case there is an additionaldifficulty, yellow, orange and red glass filters themselves fluorescewhen exposed to UV radiation. The solution is to use a UV-blockingfilter that either reflects UV-radiation in frontof the glass filters ([@fig-uvblocker-spct]). These UV-blocking filters can be, as discussed above,dichroic interference filters or filtersbased on a different pigment than the metal ions used in solid glass filters.Kodak publications from thefilm era recommend using a Kodak Wratten 2A or 2B filter in front ofother barrier filters. Tiffen still sells filters very similar to theKodak ones. In these filter a coloured light-absorbing film, in most cases usinggelatine as medium for the pigment, is encased between two layers ofclear optical glass. These filters used to be also available as baregelatine films. The Zeiss UV T\* filter is based oninterference and reflects radiation below 405 nm, it is AR coatedand less affected by reflections at the transmitted wavelengths than theTiffen. The Zeiss T\* filter prevents the fluorescence of otherglass filters as effectively or better than the Tiffen Haze 2A, but it israther expensive. The Firecrest UV400 multi-coated filter is nearly as goodas the Zeiss, but significantly cheaper. In actual use, these threefilters work well.```{r}#| label: fig-uvblocker-spct#| fig-cap: Two UV-blocking filters that do not fluoresce when exposed to UV-radiation. A Tiffen Haze 2A, equivalent to Kodak Wratten 2A and Zeiss UV T-* filter. Only the Zeiss filter has an anti-reflection coating.autoplot(filters.mspct[c("Tiffen_Haze_2A_2.6mm_52mm", "Zeiss_UV_Tstar_2.0mm_52mm")],geom ="spct", idfactor ="Filter", annotations =list(c("-", "peaks"), c("+", "wls")))```In some cases we may want to photograph fluorescence only of some colours. Highquality, dichroic band-pass filters would be perfect for this, but at reasonablylarge diameters, are very expensive. These filters are the ones used to separatefluorescence from different dyes in microscopy. If the aim is to detect allflorescence at wavelength longer than a target, it is possible to use a long-passfilter. If this is a ionic glass filter, mounted behind a UV-blocking filter([@fig-filters-VIS-pass]).```{r}#| label: fig-filters-VIS-pass#| fig-cap: Spectral transmittance of some long-pass VIS absorptive ionic glass filters. All these filters have anti-reflection multi-coating.autoplot(filters.mspct[c("Hoya_Y_(K2)_HMC_2.3mm_52mm","Heliopan_Yellow_5_SH_PMC_2.3mm_52mm", "Heliopan_Orange_22_SH_PMC_2.2mm_52mm","Heliopan_Red_25_SH_PMC_2.2mm_30.5mm","Heliopan_RG665_2.3mm_46mm")],geom ="line", idfactor ="Filter", annotations =list(c("-", "peaks"), c("+", "wls")))```With a camera modified to "see" a broader spectrum by replacement of thesensor filter, it is possible to photograph NIR fluorescence by blockingvisible fluorescence with long-pass barrier filters([@fig-filters-NIR-pass]). Compared to an unmodified camera leaks of NIRfrom the excitation radiation source become an even bigger problem. Also NIR long-pass filters require "protection" from UV to prevent theirfluorescence.::: callout-note# Fluorescence of glass filtersYellow, orange, red and NIR long-pass glass filters fluoresce whenexposed to UV-A radiation, most strongly the yellow filters ([@fig-filters-fluorescence]). Yellowfilters do emit a very strong yellow fluorescence. The fluorescence oforange and specially red filters is not so intense. Thefluorescence of NIR filters is mainly NIR and not visible to a normalcamera or the naked eye. Using a full-spectrum converted camera and along-pass filter on the lens, we can see that all four glass filtersfluoresce in the IR. ::: {#fig-filters-fluorescence layout-ncol="2"}{fig-alt="Four long-pass glass filters, yellow, orange, red and NIR"group="filters"}{fig-alt="Four filters glowing to varying extents."group="filters"}{fig-alt="Four filters glowing to varying extents."group="filters"}{fig-alt="Four filters glowing to varying extents."group="filters"}Photographs of four long-pass glass filters under white visible andultraviolet-A radiation. The background is in all photographs the sameslab of white PTFE (Teflon), but in B--D the filters used on the cameralens blocked the UV-A radiation used to excite fluorescence. Clockwisefrom the top left corner: Hoya Red 25A (603 nm), Hoya R72 (716nm),Heliopan Orange 22 (568 nm) and Heliopan Yellow 5 (466 nm), all ARcoated. Light sources: Sunwayfoto FL96 at 5500K LED fill light andJaxman Uc1 flashlight. Cameras A, B: Olympus OM-1; C, D: Olympus E-M1full-spectrum converted. For visible fluorescence barrier filter:Firecrest UV400 + Tiffen 2A, and for NIR fluorescence Firecrest UV400 +C:Zomei IR680 or D:Haida IR720 (726 nm) + Tiffen 2A. Lens: M.Zuiko 25mm 1:1.2 Pro. Photographs white-balanced and edited in Capture One Proversion 16.5.:::The spectra of the fluorescence emission have one or two peaks depending on the filter, and these peaks are at longer wavelengths than those of the radiation used for excitation ([@fig-filters-fluor-spectra]).```{r}#| label: fig-filters-fluor-spectra#| fig-cap: Fluorescence spectra for selected photography filters under excitation from 900 umol m-2 s-1 UV-A1 radiation with peak at 366 nm. A Convoy S2+ flashlight with 3W Nichia LED and a VIS-blocking filter was used as light source (spectrum in [@fig-convoy-spct]). The text header of each panel indicates supplier, filter type and filter thickness.#| fig-asp: 1.5names(filters_UVIVIF.mspct) <-gsub("_|52mm$", " ", names(filters_UVIVIF.mspct))autoplot(filters_UVIVIF.mspct, facets =2, w.band =NULL, annotations =c("-", "labels"),span =21)```To control fluorescence from these filters an additional non-fluorescentUV-reflecting or absorbing filter must be stacked in front of them. Examples ofeffective UV-blocking filters are Zeiss UV*, Firecrest UV400 and Tiffen Haze 2A.Fluorescence can be also excited by blue light, so UV-blocking filters do notalways prevent fluorescence from yellow, orange and red glass filters.:::```{r}#| label: fig-filters-NIR-pass#| fig-cap: Spectral transmittance of some long-pass NIR filters. These are all glass filters, of which the Hoya R72 is anti-reflection coated.autoplot(filters.mspct[c("Heliopan_RG695_2.2mm_52mm", "Heliopan_RG780_2.3mm_52mm","Hoya_R72_2.4mm_52mm","Zomei_IR850_2.1mm_52mm")],geom ="line", idfactor ="Filter", annotations =list(c("-", "peaks"), c("+", "wls")))```In principle band-pass filters can also be used, although this is aspecialized situation. Of band-pass filters, the only sold forphotography ready mounted on rings are the UV-IR cut filters. The actualwavelength boundaries vary rather broadly among brands and types([@fig-uvir-cut-filters]). In AliExpress and eBay sometimes unusual onesare offered.```{r, fig.height=my.fig.heigth * 3}#| label: fig-uvir-cut-filters#| fig-cap: Transmittance spectra of an assortment of UVIR-cut filters sold for photography and video.autoplot(filters.mspct[c("Firecrest_UVIR_Cut_0.96mm_52mm", "Heliopan_UVIR_CUT_Digital_2.2mm_52mm","Rocolax_UVIR_Cut_445nm_650nm_1.1mm_52mm","Rocolax_UVIR_Cut_PRO_HD_(W)_1.1mm_52mm","Fotga_UVIR_CUT_0.54mm_52mm")],geom ="line", idfactor ="Filter", facets =1,annotations =list(c("-", "peaks", "summaries"), c("+", "wls")))```Filters can be stacked, and this is frequently necessary with UV-passfilters that also transmit some NIR radiation.A huge selection of band-pass filters mounted on suitable rings areavailable from suppliers of equipment for industrial machine vision. Forexample, [MIDOPT](https://midopt.com/) (distributed in Europe by[Stemmer Imaging](https://www.stemmer-imaging.com/)), is a well knownsupplier, but such filters tend to be more expensive than those sold forphotography. Another reliable supplier with a large variety of filters is[Thorlabs](https://www.thorlabs.de), including filters specifically made for fluorescence imaging in microscopy, but mainly in smaller sizes. Square filters in a 2" $\times$ 2" or 50 mm $\times$ 50 mmsize are common in the catalogues of optics suppliers like [Heboglass](https://www.hebo-glass.com/), [UQGOptics](https://www.uqgoptics.com/), and some local optics shops, e.g.,[Teknofokus](http://www.teknofokus.fi/) in Finland.## Cameras and camera settingsNormal cameras can be used for photographing the visible fluorescence.They are preferable to full-spectrum converted cameras as the built-insensor filter help block unwanted UV and IR radiation. To photographNIR fluorescence, an IR or full-spectrum-converted camera is needed.A key requirement is that the camera supports long exposure times, i.e.,that it has at least a Bulb setting for shutter speed as exposures mayextend to minutes. Cameras that can display the live image as is"builds" during a long exposure are extremely convenient. InOlympus/OM-System cameras this mode is called and is what I always usefor UVIVF photography in the field at night. In the case of close-up andmacro photography it is possible to use exposure times that are shorterand to illuminate the whole view evenly.A lens with a large aperture helps, but is not a must as long assubjects do not move. Any good lens is suitable as the UV radiation isused to excite the fluorescence and we block it with a filter before itenters the lens. Thus, it does not require the especial lenses andmodified cameras that are required for reflected UV photography.::: callout-caution# White balancing fluorescence photographsObjectively white balancing a photograph of fluorescence is nearly impossible. After trying other approaches I have mostly settled into editing the colour in these photographs to "look right", i.e., matching my recollection of how the subject looked like when I photographed it. However, in some cases I use just a daylight balance and alternatively edit the colour to remove intense casts so as to increase the apparent colour range.:::## Close-up photography of fluorescenceCombining the close-up or macro-photography with ultraviolet-inducedvisible fluorescence in the field can be difficult as exposure times arelong and focus stacking is frequently needed for subjects with relief. Afew examples from last night can be seen in [@fig-dead-branches].::: {#fig-dead-branches layout-ncol="4"}{fig-alt="Four long-pass glass filters, yellow, orange, red and NIR"group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}{fig-alt="Four filters glowing to varying extents."group="filters"}{fig-alt="Four filters glowing to varying extents."group="filters"}Photographs of four dead tree branches, covered with lichens and mosses.The background is in all photographs the same black velvet cloth. Theedge of a slab of white PTFE (Teflon) is visible on the top-right cornerof the photographs. In, B, D, F, H, and J the filters used on the cameralens blocked the UV-A radiation used to excite fluorescence. Lightsources: Sunwayfoto FL96 at 5500K LED fill light and Jaxman Uc1flashlight. Camera: Olympus OM-1. For visible fluorescence barrierfilter: Firecrest UV400 + Tiffen 2A. Lens: A--J M.Zuiko 25 mm 1:1.2 Pro,K-L M.Zuiko 90 mm 1:3.5 Macro Pro. Photographs white-balanced and editedin Capture One Pro version 16.5. Focus-stacks of 20 or 80 photographswere merged in Helicon Focus 8.2.18 Pro to increase the depth of field.:::With suitable filters and lenses we can even attempt macro-photography of UV-A-induced chlorophyll fluorescence with a modified mirrorless camera ([@fig-uvinirf]).::: {#fig-uvinirf layout-ncol="2" width="50%}{fig-alt="Four long-pass glass filters, yellow, orange, red and NIR"group="branches"}{fig-alt="Four filters glowing to varying extents."group="branches"}Moss on a dead branch. Illuminated with A: white light or B: UV-A radiation. Only difference with images in [@fig-dead-branches] in the approach was adding a long-pass filter with a cut-in at 650 nm (HB650). Using a filter with cut-in at 695 nm (Heliopan RG695) resulted in a similar image but required a longer exposure time. Camera: full-spectrum converted Olympus E-M1 camera. See legend to [@fig-dead-branches] for other details.:::## ConclusionPhotographing UV-A-induced visible fluorescence is interesting and lessdemanding in equipment and technique than photography in reflected UV radiation.An unmodified digital camera can be used and long pass filters are in generalcheaper and more common than UV short-pass filters that block visible andinfrared radiation effectively enough. UV-A as excitation is convenient and easyto work with, blue light could be used instead if we are interested influorescence at longer wavelengths or for safety reasons in teaching situations.Using UV-B radiation to induce UV-A and violet fluorescence is in principlepossible but more demanding in practice. Photographing visible-light induced near infrared fluorescence is more demanding in equipment,but opens interesting possibilities. Could we with comparatively cheap equipmentuse photography to qualitatively assess changes epidermal UV transmittance undercontrolled conditions?## Other sources of information and helpThe [_UltravioletPhotography_ discussion group](https://www.ultravioletphotography.com/) is a very friendly and knowledgeable group of people from around the world with an interest in UV, NIR, and UV-induced fluorescence.The web sites of Adrian Davies, author of three of the books listed below, include galleries of photographs, including some with photographs of UV-A induced fluorescence. See [Imaging the Invisible](https://www.imagingtheinvisible.com/gallery_714121.html) and [Adrian Davies imaging](https://www.adriandaviesimaging.com).## Further reading